Headquarters Daily report MAY 25, 1994 *************************************************************************** REPORT NEGATIVE NO INPUT ATTACHED INPUT RECEIVED RECEIVED HEADQUARTERS X REGION I X REGION II X REGION III X REGION IV X *************************************************************************** PRIORITY ATTENTION REQUIRED MORNING REPORT - HEADQUARTERS MAY 25, 1994 MR Number: H-94-0052 NRR DAILY REPORT ITEM GENERIC COMMUNICATIONS BRANCH/EVENTS ASSESSMENT BRANCH DIVISION OF OPERATING REACTOR SUPPORT OFFICE OF NUCLEAR REACTOR REGULATION Subject: N/A NRC Information Notice 94-37, "Misadministration Caused by a Bent Interstitial Needle During Brachytherapy Procedure, to be issued May 27, 1994. NRC is alerting licensees of an incident involving an interstitial needle bent inside the patient's body during a high-dose-rate procedure with an Omnitron 2000 brachytherapy system. The bend in the Omnitron interstitial needle, through which the radioactive source travels, prevented the source from retracting beyond the point of the bend. This resulted in the actual radiation dose received by the patient exceeding the prescribed dose by approximately 75 percent. Technical contacts: Hector Bermudez, RII (404) 331-7880 James Smith, NMSS (301) 415-7904 ************************************************************************* NRC Information Notice 94-38, "Results of a Special NRC Inspection at Dresden Nuclear Power Station Unit 1 Following a Rupture of Service Water Inside Containment," to be issued May 27, 1994. The NRC is issuing this information notice to inform addressees of the results of a special NRC inspection at Dresden Nuclear Power Station Unit 1 (Dresden 1) after a rupture of the service water system occurred inside the containment. Technical contacts: James McCormick-Barger, RIII (708) 829-9872 Richard Dudley, NRR (301) 504-1116 Larry Bell, NMSS (301) 504-2171